The uncertainty calculation in the experiment of measurement of sodium wavelength with michelson interferometers

Expand
  • Department of Science, Shenyang Aerospace University, Shenyang 110136

Received date: 2013-03-19

Abstract

In this paper, the measurement of the sodium light wavelength by Michelson interferometer has been taken as an example to consider the practical significance of single measurement by analysing the uncertainty, and taking the position value of moving mirror as a direct measurement target, and the change value of moving mirror as an indirect measurement target which depends on the position value of moving mirror when the interference fringes change fifty times.In comparison with the usual processing method, it was observed that even if the displacement distance of moving mirror was taken as a direct measurement target, the instrument error could not be simply taken as the total uncertainty of B component.Taking the above two kinds of circumstances into account, the analysis of estimating uncertainty in processing the experimental data through gradual deduction method is suggested.

Cite this article

PAN Zhe-feng, SUN Jing-chao . The uncertainty calculation in the experiment of measurement of sodium wavelength with michelson interferometers[J]. Journal of Shenyang Aerospace University, 2013 , 30(4) : 89 -90 . DOI: 10.3969/j.issn.2095-1248.2013.04.018